Comparison of simultaneous multi-slice readout-segmented EPI and conventional single-shot EPI for diffusion tensor imaging of the ulnar nerve
Comparison of simultaneous multi-slice readout-segmented EPI and conventional single-shot EPI for diffusion tensor imaging of the ulnar nerve
复制标题
DOI:
10.1016/j.heliyon.2018.e00853
复制
发表时间:
2018-10-01
期刊:
影响因子:
4
通讯作者:
Filli, Lukas
中科院分区:
文献类型:
--
作者:
Ho, Michael;Becker, Anton;Filli, Lukas
Purpose: To compare conventional single-shot echo planar imaging (ss-EPI) and simultaneous multi-slice (SMS) readout-segmented EPI (rs-EPI) for magnetic resonance diffusion tensor imaging (DTI) of the ulnar nerve.Materials and methods: This study was approved by the local ethics committee. Ten healthy volunteers (mean age 30.4 +/- 4.01 years; range 25-36 years) underwent 3T DTI of the ulnar nerve at the level of the cubital tunnel. DTI was performed based on ss-EPI as well as SMS rs-EPI sequences. Signal-to-noise ratio (SNR), image quality, and DTI parameters in the ulnar nerve (fractional anisotropy, FA; mean diffusivity, MD) were compared between the two sequences by two independent radiologists.Results: Acquisition time was 5:12 min for ss-EPI and 5:18 min for SMS rs-EPI. Between the two sequences, no significant differences were found for derived DTI measures FA (p = 0.11) and MD values (p = 0.93). Compared to conventional ss-EPI, SMS rs-EPI yielded significantly less ghosting artifacts (p = 0.04) but inferior nerve depiction (p = 0.001) and worse overall image quality (p = 0.008).Conclusion: SMS rs-EPI is not advantageous over ss-EPI in DTI of the ulnar nerve at the level of the cubital tunnel.