Measurements and detection of abdominal aortic aneurysm growth: Accuracy and reproducibility of a segmentation software
Measurements and detection of abdominal aortic aneurysm growth: Accuracy and reproducibility of a segmentation software
复制标题
DOI:
10.1016/j.ejrad.2011.04.044
复制
发表时间:
2012-08-01
影响因子:
3.3
通讯作者:
Soulez, Gilles
中科院分区:
文献类型:
--
作者:
Kauffmann, Claude;Tang, An;Soulez, Gilles
Purpose: To validate the reproducibility and accuracy of a software dedicated to measure abdominal aortic aneurysm (AAA) diameter, volume and growth over time.Materials and methods: A software enabling AAA segmentation, diameter and volume measurement on computed tomography angiography (CTA) was tested. Validation was conducted in 28 patients with an AAA having 2 consecutive CTA examinations. The segmentation was performed twice by a senior radiologist and once by 3 medical students on all 56 CTAs. Intra and inter-observer reproducibility of D-max and volumes values were calculated by intraclass correlation coefficient (ICC). Systematic errors were evaluated by Bland-Altman analysis. Differences in D-max and volume growth were compared with paired Student's t-tests.Results: Mean D-max and volume were 49.6 +/- 6.2 mm and 117.2 +/- 36.2 ml for baseline and 53.6 +/- 7.9 mm and 139.6 +/- 56.3 ml for follow-up studies. Volume growth (17.3%) was higher than D-max progression (8.0%) between baseline and follow-up examinations (p < .0001). For the senior radiologist, intra-observer ICC of D-max and volume measurements were respectively estimated at 0.997 (>= 0.991) and 1.000 (>= 0.999). Overall inter-observer ICC of D-max and volume measurements were respectively estimated at 0.995 (0.990-0.997) and 0.999 (>0.999). Bland-Altman analysis showed excellent inter-reader agreement with a repeatability coefficient