YAP-(S)PET small animal scanner: Quantitative results
YAP-(S)PET small animal scanner: Quantitative results
复制标题
DOI:
10.1109/tns.2003.817970
复制
发表时间:
2003-10-01
影响因子:
1.8
通讯作者:
Del Guerra, A
中科院分区:
文献类型:
--
作者:
Di Domenico, G;Zavattini, G;Del Guerra, A
The University of Ferrara YAP-(S)PET scanner is currently being employed in small animal SPECT studies, with Tc-99 m labeled radiotracers, with the goal of obtaining quantitative activity measurements from region of interest (ROI) analysis of reconstructed images of rats. The measurements will be compared directly with traditional ex vivo measurement of activity with gamma counters. To achieve this goal, the scanner was calibrated relative to a standard Isodose dose calibrator which was calibrated with various certified activity sources. The calibration factor K was defined as the ratio of the activity measured with the Isodose (MBq) and the image count rate (cps). We find K = 0.1346 +/- 0.0008 MBq/cps, with good linearity over a wide range of activities (9-88 MBq). With the calibrated scanner we compared results of activity measurements from images of whole-rat heart acquisitions versus excised hearts. We found good agreement between the two measurements.