Principles of stopped-flow sequential injection analysis and its application to the kinetic determination of traces of a proteolytic enzyme.
Principles of stopped-flow sequential injection analysis and its application to the kinetic determination of traces of a proteolytic enzyme.
复制标题
停流连续注射分析原理及其在蛋白水解酶痕量动力学测定中的应用。
DOI:
10.1021/ac00017a006
复制
发表时间:
1991
影响因子:
7.4
通讯作者:
Gübeli,T
中科院分区:
文献类型:
--
作者:
Ruzicka,J;Gübeli,T
The concept of sequential Injection analysis Is described, and an application of this novel technique to an assay of traces of a proteolytic enzyme Is demonstrated. The automated method Is based on a stopped-flow procedure with fluores-cence detection. Instead of using the conventional constant flow rate, a sinusoidal flow pattern, as generated by a robust cam-driven piston pump, has been applied. The validation of the novel flow Injection system using a nonreactlve fluorescent compound yielded highly reproducible results over a concentration range of 3 orders of magnitude. The calibration graph for the enzyme activities In the range 5-75 pu mL~ 1 could be fitted a second-order polynomial. The detection Rmlt Is 2.5/uu mL~\which corresponds to 7.2 ng mL~ 1 pure active enzyme.