Real-time SPARSE-SENSE cine MR imaging in atrial fibrillation: a feasibility study
Real-time SPARSE-SENSE cine MR imaging in atrial fibrillation: a feasibility study
复制标题
DOI:
10.1177/0284185116681037
复制
发表时间:
2017-08-01
期刊:
影响因子:
1.3
通讯作者:
Nassenstein, Kai
中科院分区:
文献类型:
--
作者:
Goebel, Juliane;Nensa, Felix;Nassenstein, Kai
Background: Cardiac magnetic resonance imaging (MRI) relies on correct ECG-gating, which is hindered in arrhythmia.Purpose: To examine whether a prototype free-breathing real-time cine sequence using SPARSE-SENSE (SPARSE) improves left ventricular quantification in atrial fibrillation.Material and Methods: On a 1.5T MR system left ventricular short-axis stacks were acquired of the SPARSE sequence and of a reference steady-state free precession (SSFP) sequence with arrhythmia rejection in 20 patients with atrial fibrillation. Two radiologists independently rated arrhythmia-caused artifact severity in both sequences using a 4-point scale. Coefficients of variation of myocardial signal intensity for both sequences were acquired. Volumetry was performed twice by one reader and once by another reader. Correlation between artifact severity and employed sequence was analyzed by modified Fisher's exact test. Coefficients of variation and volumetric data were compared by paired t-test and intraclass correlation.Results: Median arrhythmia-caused artifact severity was 2 in both readers for SSFP and 0 (reader 1)/1 (reader 2) for SPARSE, being significantly lower in SPARSE (P