Abdomen:: Diffusion-weighted MR imaging with pulse-triggered single-shot sequences

Abdomen:: Diffusion-weighted MR imaging with pulse-triggered single-shot sequences
复制标题

DOI:
10.1148/radiol.2241011117
复制
发表时间:
2002-07-01
期刊:
影响因子:
19.7
通讯作者:
Schild, HH
Schild, HH
中科院分区:
医学1区
文献类型:
--
作者:
Mürtz, P;Flacke, S;Schild, HH

文献摘要

被引文献

相似文献

在12名健康志愿者的腹部进行磁共振(MR)扩散测量,通过使用扩散加权单激发序列无和脉冲触发不同的触发延迟。脉冲触发至心脏舒张期可减少扩散加权MR图像上的运动伪影,并显着提高腹部器官表观扩散系数(ADC)测量的准确性和可重复性。(C)RSNA,2002年。
Magnetic resonance (MR) diffusion measurements of the abdomen were performed in 12 healthy volunteers by using a diffusion-weighted single-shot sequence both without and with pulse triggering for different trigger delays. Pulse triggering to the diastolic heart phase led to reduced motion artifacts on the diffusion-weighted MR images and to significantly improved accuracy and reproducibility of measurements of the apparent diffusion coefficients, or ADCs, of abdominal organs. (C) RSNA, 2002.