PERFORMANCE EVALUATION OF DIAGNOSTIC RADIOLOGY DOSIMETERS IN CLINICAL AND CALIBRATION X-RAY BEAMS
PERFORMANCE EVALUATION OF DIAGNOSTIC RADIOLOGY DOSIMETERS IN CLINICAL AND CALIBRATION X-RAY BEAMS
复制标题
DOI:
10.1097/hp.0b013e3181d18d45
复制
发表时间:
2010-05-01
期刊:
影响因子:
2.2
通讯作者:
Manetou, Aggeliki
中科院分区:
文献类型:
--
作者:
Hourdakis, Costantine John;Boziari, Argyro;Manetou, Aggeliki
Diagnostic radiology dosimeters should comply with International Electrotechnical Commission (IEC) 61674 standard in order to perform measurements with sufficient accuracy and reliability. The calibration of a dosimeter is performed under, and pertains to, reference conditions. However, in most cases, dosimeters are used for clinical measurements under non-reference conditions. The performance, in terms of accuracy of dose measurements, of six commercial diagnostic radiology dosimeters was tested at reference calibration and at clinical non-reference conditions. The results showed that all dosimeters being tested exhibited limits of variation within the +/- 5% IEC limits. Depending on the detector's physical and operational properties, the dosimeters' energy dependence of response values varied from -4.7% to +4.2%. To address this variation of response, calibration at three radiation qualities (RQR 3, RQR 5, and RQR 9), at least, is recommended. Different irradiation conditions such as air kerma rate, x-ray tube design, x-ray system, and dosimeter operational modes affect the dosimeters' response by less than 3%. A dosimeter that complies with IEC standards and operates according to its specifications could be used at typical clinical irradiation conditions taking into account only corrections for the energy dependence of response. In this case, the error in dose accuracy is expected to be less than 3%. Health Phys. 98(5): 704-716; 2010