Comparison of volatile sulfur compound concentrations measured with a sulfide detector vs. gas chromatography
Comparison of volatile sulfur compound concentrations measured with a sulfide detector vs. gas chromatography
复制标题
DOI:
10.1177/154405910208100211
复制
发表时间:
2002-02-01
影响因子:
7.6
通讯作者:
Levitt, MD
中科院分区:
文献类型:
--
作者:
Furne, J;Majerus, G;Levitt, MD
The accuracy of the Halimeter(R), an inexpensive, simple instrument that measures total breath volatile sulfur compounds (VSCs), has not been adequately tested. We compared Halimeter(R) measurements with those obtained with a specific and sensitive gas chromatographic (GC) technique. The Halimetere, gave different, biexponential responses to a constant concentration of different VSCs: The relative response rate and sensitivity were hydrogen sulfide > methyl mercaptan > dimethylsulfide. The transient peak VSC concentration of oral samples was reached long before the sulfide detector fully responded. The GC measurement of initial total VSCs in breath samples was 2.7 +/- 0.48 times greater than the peak concentration of the Halimeter(R). However, the plateau phase measurement of the Halimeter(R) was 25% greater than that of GC. While GC and Halimeter(R) measurements positively correlated, appreciable differences were observed. In studies where relatively precise VSC measurements are required, GC is the preferable technique.