Non-contact respiratory rate measurement validation for hospitalized patients.

Non-contact respiratory rate measurement validation for hospitalized patients.
复制标题

DOI:
10.1109/iembs.2009.5332635
复制
发表时间:
2009
期刊:
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
影响因子:
--
通讯作者:
Boric-Lubecke O
Boric-Lubecke O
中科院分区:
其他
文献类型:
--
作者:
Droitcour AD;Seto TB;Park BK;Yamada S;Vergara A;El Hourani C;Shing T;Yuen A;Lubecke VM;Boric-Lubecke O

文献摘要

被引文献

相似文献

This paper presents the first clinical results for validating the accuracy of respiratory rate obtained for hospitalized patients using a non-contact, low power 2.4 GHz Doppler radar system. Twenty-four patients were measured in this study. The respiratory rate accuracy was benchmarked against the respiratory rate obtained using Welch Allyn Propaq Encore model 242, the Embla Embletta system with Universal XactTrace respiratory effort sensor and Somnologica for Embletta software, and by counting chest excursions. The 95% limits of agreement between the Doppler radar and reference measurements fall within +/−5 breaths per minute.