Practical test of a point-source integrating cavity absorption meter: the performance of different collector assemblies.
Practical test of a point-source integrating cavity absorption meter: the performance of different collector assemblies.
复制标题
点源积分腔吸收计的实际测试:不同收集器组件的性能。
作者:
R. Röttgers;W. Schönfeld;P. Kipp;R. Doerffer
A prototype point-source integrating cavity absorption meter (PSICAM) is presented and compared with spectrophotometric absorption measurements. Different light collector assemblies of the PSICAM were tested regarding their capability to determine the absorption of water constituents accurately over a wide range of concentrations and scattering properties. The PSICAM setup with a radiance-type sensor showed the best performance. It was compared with a photometric absorption determination using nonscattering dye solutions. The mean difference between both methods was less than 2.4% in the spectral range of 400-700 nm. The absorption determination with the PSICAM, when equipped with a radiance sensor as a light collector, was only little affected by scattering and temperature. We conclude that the PSICAM can be used to determine the absorption of natural seawater samples at ambient temperatures.