Patient-induced susceptibility effect on geometric distortion of clinical brain MRI for radiation treatment planning on a 3T scanner
Patient-induced susceptibility effect on geometric distortion of clinical brain MRI for radiation treatment planning on a 3T scanner
复制标题
DOI:
10.1088/0031-9155/58/3/465
复制
发表时间:
2013-02-07
影响因子:
3.5
通讯作者:
Cao, Y.
中科院分区:
文献类型:
--
作者:
Wang, H.;Balter, J.;Cao, Y.
Concerns about the geometric accuracy of MRI in radiation therapy (RT) have been present since its invention. Although modern scanners typically have system levels of geometric accuracy that meet requirements of RT, subject-specific distortion is variable, and methods to in vivo assess and control patient-induced geometric distortion are not yet resolved. This study investigated the nature and magnitude of the subject-induced susceptibility effect on geometric distortions in clinical brain MRI, and tested the feasibility of in vivo quality control using field inhomogeneity mapping. For 19 consecutive patients scanned on a dedicated 3T MR scanner, B0 field inhomogeneity maps were acquired and analyzed to determine subject-induced distortions. For 3DT1 weighted images frequency-encoded with a bandwidth of 180 Hz/pixel, 86.9% of the estimated displacements were