The Reliability and Variation of Acetabular Component Anteversion Measurements From Cross-Table Lateral Radiographs
The Reliability and Variation of Acetabular Component Anteversion Measurements From Cross-Table Lateral Radiographs
复制标题
DOI:
10.1016/j.arth.2011.03.039
复制
发表时间:
2011-09-01
影响因子:
3.5
通讯作者:
Barrack, Robert L.
中科院分区:
文献类型:
--
作者:
Nunley, Ryan M.;Keeney, James A.;Barrack, Robert L.
Although cross-table lateral (CL) radiographs are frequently used to assess acetabular component anteversion, the reliability of this method has not been established. We compared serial CL radiographs with computed tomography (CT) scans for 98 total hip arthroplasty patients (119 hips) undergoing surveillance of primary or revision total hip arthroplasty. Acetabular anteversion averaged 26.1 degrees (range, -2 degrees to 48.3 degrees) on CL imaging and 28.8 degrees (range, -7 degrees to 54 degrees) on CT scan. There was a strong correlation between anteversion determined from CT scans and serial CL images. However, variation on serial CL studies exceeded 10 degrees for 20% of patients. Although CL imaging provides acceptable assessment of general component position, it has limited use for precise analysis in research, outcome reporting, or determination of cause of implant failure.