Intraoperative handheld optical coherence tomography forward-viewing probe: physical performance and preliminary animal imaging.
Intraoperative handheld optical coherence tomography forward-viewing probe: physical performance and preliminary animal imaging.
复制标题
DOI:
10.1364/boe.3.001404
复制
发表时间:
2012-06-01
影响因子:
3.4
通讯作者:
Yang VX
中科院分区:
文献类型:
--
作者:
Sun C;Lee KK;Vuong B;Cusimano MD;Brukson A;Mauro A;Munce N;Courtney BK;Standish BA;Yang VX
A prototype intraoperative hand-held optical coherence tomography (OCT) imaging probe was developed to provide micron resolution cross-sectional images of subsurface tissue during open surgery. This new ergonomic probe was designed based on electrostatically driven optical fibers, and packaged into a catheter probe in the form factor of clinically accepted Bayonet shaped neurosurgical probes. Optical properties of the probe were measured to have a ~20 μm spot size, 5 mm working distance and 4 mm field of view. Feasibility of this probe for structural and Doppler shift imaging was tested on porcine femoral blood vessel imaging.