Advancing the Hexapole Cs-Corrector for the Scanning Transmission Electron Microscope
Advancing the Hexapole Cs-Corrector for the Scanning Transmission Electron Microscope
复制标题
DOI:
10.1017/s1431927606060600
复制
发表时间:
2006-10
影响因子:
2.8
通讯作者:
H. Müller;S. Uhlemann;P. Hartel;M. Haider
中科院分区:
文献类型:
--
作者:
H. Müller;S. Uhlemann;P. Hartel;M. Haider
Aberration correctors using hexapole fields have proven useful to correct for the spherical aberration in electron microscopy. We investigate the limits of the present design for the hexapole corrector with respect to minimum probe size for the scanning transmission electron microscope and discuss several ways in which the design could be improved by rather small and incremental design changes for the next generation of advanced probe-forming systems equipped with a gun monochromator.