Operation of a 130-MHz/9-mm Compact HTS Annulus Magnet With a Micro-NMR Probe.
Operation of a 130-MHz/9-mm Compact HTS Annulus Magnet With a Micro-NMR Probe.
复制标题
使用微型 NMR 探头操作 130 MHz/9 mm 紧凑型 HTS 环形磁铁。
DOI:
10.1109/tasc.2014.2367658
复制
发表时间:
2015
期刊:
影响因子:
--
通讯作者:
Iwasa,Yukikazu
中科院分区:
文献类型:
--
作者:
Hahn,Seungyong;Kim,Youngjae;Song,Jungbin;Voccio,JohnP;Chu,Yong;Bascuñán,Juan;Tomita,Masaru;Iwasa,Yukikazu
In this paper, we report final operation results of our compact annulus NMR magnet, named YP2800, with a homemade micro-NMR probe in a bath of liquid helium at 4.2 K. YP2800 comprises of a stack of 2800 YBCO “plate annuli,” 0.08 mm thick, either 46 mm or 40 mm square, each having a 26-mm hole machined at the center. By the field-cooling technique, YP2800 was energized at 130 MHz (3.05 T); an overall peak-to-peak homogeneity of 487 ppm withinwas measured at a moment when a field drift of 11 ppm/h was reached in three days after field cooling. Due to a small (9.2 mm) bore size, no commercial probes could fit into the bore; an 8.5-mm micro-NMR probe was designed and constructed. Following a general description of YP2800 and design construction details of the micro probe, this paper presents NMR signals captured by the probe for a dimethyl sulfoxide sample of4.4 and 5 mm long at a base frequency of 130 MHz with a half-peak width of 60 kHz; the corresponding frequency impurity of 461 ppm is chiefly due to a spatial field error, i.e., 487 ppm in the target space.