Dynamic calibration method of temperature sensor based on Quasi-delta pulse temperature excitation
Dynamic calibration method of temperature sensor based on Quasi-delta pulse temperature excitation
复制标题
基于准δ脉冲温度激励的温度传感器动态校准方法
DOI:
10.1002/mop.30944
复制
发表时间:
2018
影响因子:
1.5
通讯作者:
Cheng Hang
中科院分区:
文献类型:
--
作者:
Li Yanfeng;Zhang Zhijie;Hao Xiaojian;Cheng Hang
Aiming at the existing problems in the dynamic calibration method of temperature sensor, Quasi‐δ pulse temperature dynamic calibration method is proposed. The Quasi‐δ pulse dynamic calibration principle was described, the characteristics of Quasi‐δ pulse function were analyzed, and the relationship among pulse width of Quasi‐δ signal, frequency range that can be calibrated, and calibration precision was obtained. Quasi‐δ pulse temperature dynamic calibration system was built using the laser as the heat source which generates the excitation temperature required for dynamic calibration. In the system, optical focusing system was designed using ellipsoidal mirror which has a better conjugate imaging property. Applying the dynamic calibration system, a K‐type armored thermocouple was calibrated dynamically, and mathematical model of the calibrated thermocouple was established by generalized least squares method based on a special whitening filter. Dynamic characteristics of the thermocouple in time domain and frequency domain were analyzed.