In situ measurements of krypton in xenon gas with a quadrupole mass spectrometer following a cold-trap at a temporarily reduced pumping speed
In situ measurements of krypton in xenon gas with a quadrupole mass spectrometer following a cold-trap at a temporarily reduced pumping speed
复制标题
DOI:
10.1088/1748-0221/8/02/p02011
复制
发表时间:
2012-12
影响因子:
1.3
通讯作者:
E. Brown;S. Rosendahl;C. Huhmann;C. Weinheimer;Hans Kettling
中科院分区:
文献类型:
--
作者:
E. Brown;S. Rosendahl;C. Huhmann;C. Weinheimer;Hans Kettling
A new method for measuring trace amounts of krypton in xenon using a cold trap with a residual gas analyzer has been developed, which achieves an increased sensitivity by temporarily reducing the pumping speed while expending a minimal amount of xenon. By partially closing a custom built butterfly valve between the measurement chamber and the turbomolecular pump, a sensitivity of 40 ppt has been reached. This method has been tested on an ultra-pure gas sample from Air Liquide with an unknown intrinsic krypton concentration, yielding a krypton concentration of 330±200 ppt.