Potential of image-guidance, gating and real-time tracking to improve accuracy in pulmonary stereotactic body radiotherapy
Potential of image-guidance, gating and real-time tracking to improve accuracy in pulmonary stereotactic body radiotherapy
复制标题
DOI:
10.1016/j.radonc.2008.08.010
复制
发表时间:
2009-06-01
影响因子:
5.7
通讯作者:
Flentje, Michael
中科院分区:
文献类型:
--
作者:
Guckenberger, Matthias;Krieger, Thomas;Flentje, Michael
Purpose: To evaluate the potential of image-guidance, gating and real-time tumor tracking to improve accuracy in pulmonary stereotactic body radiotherapy (SBRT).Materials and methods: Safety margins for compensation of inter- and intra-fractional uncertainties of the target position were calculated based on SBRT treatments of 43 patients with pre- and post-treatment cone-beam CT imaging. Safety margins for compensation of breathing motion were evaluated for 17 pulmonary tumors using respiratory correlated CT, model-based segmentation of 4D-CT images and voxel-based dose accumulation; the target in the mid-ventilation position was the reference.Results: Because of large inter-fractional base-line shifts of the tumor, stereotactic patient positioning and image-guidance based on the bony anatomy required safety margins of 12 mm and 9 mm, respectively. Four-dimensional image-guidance targeting the tumor itself and intra-fractional tumor tracking reduced margins to