Performance improvement of Si(Li) Peltier cooled detectors
复制标题
硅(锂)帕尔贴冷却探测器的性能改进
DOI:
--
复制
发表时间:
2001
期刊:
影响因子:
--
通讯作者:
V. A. Zalinkevich
中科院分区:
文献类型:
--
作者:
A. Sokolov;V. Gostilo;A. Loupilov;V. A. Zalinkevich
Spectrometric performance improvement of Si(Li) Peltier cooled detectors obtained by improvement of electronics performance, development of the small-sized vacuum chamber and effective 5-channel Peltier cooler, are providing temperature difference of 132/spl deg/C at sensitivity to heat loading less than 15 mWt//spl deg/C. It allows to fabricate circular Si(Li) Peltier cooled detectors with thickness of 4-4.5 mm and sensitive surface from 12 till 100 mm/sup 2/. Their energy resolution on energy 5.9 keV was from 158 till 2 40 eV depending on the sensitive surface area. And energy resolution on 59.6 keV for these detectors was from 410 till 460 eV. Rectangular detectors with thickness 4-4.5 mm and sensitive surface 60 mm/sup 2/ had energy resolution 255 and 450 eV on energies 5.9 and 59.6 keV correspondingly.