CALIBRATION OF [C-14] PLASTIC STANDARDS FOR QUANTITATIVE AUTORADIOGRAPHY OF [I-125]LABELED LIGANDS WITH AMERSHAM HYPERFILM BETA-MAX
CALIBRATION OF [C-14] PLASTIC STANDARDS FOR QUANTITATIVE AUTORADIOGRAPHY OF [I-125]LABELED LIGANDS WITH AMERSHAM HYPERFILM BETA-MAX
复制标题
DOI:
10.1016/0304-3940(89)90350-9
复制
发表时间:
1989-09-25
影响因子:
2.5
通讯作者:
WIMPY, TH
中科院分区:
文献类型:
--
作者:
BASKIN, DG;WIMPY, TH
The densitometric response of Hyperfilm .beta.-max (HFBM) to [14C] plastic standards was calibrated to tissue-equivalent concentrations of [125I]. Plastic sections with standard concentration of [125C] were apposed to HFBM for 1, 3, 5, and 9 days, with liver and muscle slices (20 .mu.m thick) labeled with [125I] and [14C] insulin. The relative optical densities (ROD = log10 [1/gray level .times. 256-1] produced by [125I]- and [14C] plastic standards were converted to equivalent tissue [125I] concentrations (dpm/mm2). The response of HFBM to the [125I]- and [14C] plastic standards was similar (P < 0.001). Standard curves of tissue [125I] (dpm/mm2) vs plastic [14C] (.mu.Ci/g) concentrations fit second order polynomials (r2 = 0.995-0.999). The results show that [14C] plastic standards are valid for measuring [125I] radioactivity in tissue slices by autoradiography with HFBM.