Initial experience with the Pascal photocoagulator: a pilot study of 75 procedures
Initial experience with the Pascal photocoagulator: a pilot study of 75 procedures
复制标题
DOI:
10.1136/bjo.2008.139568
复制
发表时间:
2008-08-01
影响因子:
4.1
通讯作者:
Stanga, P. E.
中科院分区:
文献类型:
--
作者:
Sanghvi, C.;McLauchlan, R.;Stanga, P. E.
Background: The Pascal is a semiautomated photocoagulator that delivers a pattern array of multiple burns in a rapid predetermined sequence with a single foot pedal depression. Each burn is reduced to 10 or 20 ms to achieve this. The authors report their early experience with this system.Methods: 75 procedures done in 60 patients divided into four groups-group A, patients undergoing panretinal photocoagulation (PRP); group B, patients undergoing focal or modified grid macular laser; group C, patients undergoing macular grid and group D, patients undergoing retinopexy-were retrospectively studied.Results: 31/34 procedures in group A, 24/26 procedures in group B, 5/7 procedures in group C and all eight patients in group D had successful outcomes. Significantly higher powers were required with the Pascal than with conventional laser (p