A method for calibrating the line-focus-beam acoustic microscopy system
A method for calibrating the line-focus-beam acoustic microscopy system
复制标题
一种线聚焦声学显微镜系统的校准方法
DOI:
10.1109/58.660152
复制
发表时间:
1998
期刊:
影响因子:
--
通讯作者:
M. Arakawa
中科院分区:
文献类型:
--
作者:
J. Kushibiki;M. Arakawa
Absolute accuracy of the line-focus-beam (LFB) acoustic microscopy system is investigated for measurements of the leaky surface acoustic wave (LSAW) velocity and attenuation, and a method of system calibration is proposed. In order to discuss the accuracy, it is necessary to introduce a standard specimen whose bulk acoustic properties, (e.g., the independent elastic constants and density) are measured with high accuracy. Single crystal substrates of gadolinium gallium garnet (GGG) are taken as standard specimens. The LSAW propagation characteristics are measured and compared with the calculated results using the measured bulk acoustic properties. Calibration is demonstrated for the system using two LFB acoustic lens devices with a cylindrical concave surface of 1-mm radius in the frequency range 100 to 300 MHz.