A comparison of the imaging performance of different PSPMTs for PET applications
A comparison of the imaging performance of different PSPMTs for PET applications
复制标题
PET 应用中不同 PSPMT 的成像性能比较
DOI:
10.1016/j.nima.2003.11.033
复制
发表时间:
2004
影响因子:
1.4
通讯作者:
A. Guerra
中科院分区:
文献类型:
--
作者:
D. Herbert;N. Belcari;M. Camarda;A. Guerra
Three different types of Position Sensitive PhotoMultiplier tube have been studied and compared in terms of their imaging performance when coupled to scintillator arrays. Such position sensitive tubes often form the basis for the coincident detection of the 511keV gamma-rays produced as a result of positron annihilation in PET. The three tubes studied here are the R2486, the R8520 and the H8500 all from Hamamatsu and span a number of years of technological development. The R2486 is a circular tube that has a photocathode of diameter 50mm and a body diameter of 76mm. The R8520 is a compact 25.7×25.7mm2square detector with a 22×22mm2active area. The H8500, also called the flat panel tube, is a large area square detector of 52×52mm2with an active area of 49×49mm2. A 40×40mm2YAP:Ce matrix composed of 2×2×25mm3pixels is used in order to assess the imaging performance of each tube at 511keV. This imaging performance is characterised here in terms of the image pixel FWHM and the peak-to-valley ratio. After the presentation of these results a discussion is given of the relative merits of each tube and its suitability for future PET applications.