Calibration of photomultiplier tubes for the fluorescence detector of tele scope array experiment using a Rayleigh scattered laser beam

Calibration of photomultiplier tubes for the fluorescence detector of tele scope array experiment using a Rayleigh scattered laser beam
复制标题

瑞利散射激光束校准望远镜阵列实验荧光探测器光电倍增管

DOI:
10.1016/j.nima.2012.03.011
复制
发表时间:
2012
期刊:
Nuclear Instruments and Method in Physics Research Section A
影响因子:
--
通讯作者:
et al.
et al.
中科院分区:
--
文献类型:
--
作者:
S.Kawana;T.Fujii(l5人中3番目);et al.

文献摘要

相似文献

我们进行了光度校准的光电倍增管(PMT)和读出电子用于新的荧光探测器的望远镜阵列(TA)实验使用瑞利散射光子从脉冲氮激光束。实验装置,测量程序,校准的结果进行了说明。校准的总系统不确定度估计为7.2%。一个额外的不确定度为3.7%的引入由校准的PMT从实验室到TA实验场所的运输。
We performed photometric calibration of the PhotoMultiplier Tube (PMT) and readout electronics used for the new fluorescence detectors of the Telescope Array (TA) experiment using Rayleigh scattered photons from a pulsed nitrogen laser beam. The experimental setup, measurement procedure, and results of calibration are described. The total systematic uncertainty of the calibration is estimated to be 7.2%. An additional uncertainty of 3.7% is introduced by the transport of the calibrated PMTs from the laboratory to the TA experimental site.