Phantom design method for high-field MRI human systems

Phantom design method for high-field MRI human systems
复制标题

DOI:
10.1002/mrm.20245
复制
发表时间:
2004-11-01
影响因子:
3.3
通讯作者:
Chen, W
Chen, W
中科院分区:
医学3区
文献类型:
--
作者:
Yang, QX;Wang, JH;Chen, W

文献摘要

被引文献

相似文献

A phantom design method suitable for high-field MRI based on the RF field wave characteristics of sample and experimental validations at 7.0 T and 3.0 T are presented. The RF field distribution in a phantom with a given RF coil system is primarily determined by the sample size relative to the wavelength inside the sample, and the ratio between the displacement and conduction currents. Experimental results demonstrate that the MR image intensity patterns associated with wave behavior in human samples at a given field strength can be reproduced with a phantom at the same or different field strengths once the dimension and penetration constant are scaled by the corresponding wavelength in the sample medium. (C) 2004 Wiley-Liss, Inc.