Comparison of phase-shifting techniques for in vivo full-range, high-speed Fourier-domain optical coherence tomography.
Comparison of phase-shifting techniques for in vivo full-range, high-speed Fourier-domain optical coherence tomography.
复制标题
DOI:
10.1117/1.3494556
复制
发表时间:
2010-09
影响因子:
3.5
通讯作者:
D. Kim;J. Werner;R. Zawadzki
中科院分区:
文献类型:
--
作者:
D. Kim;J. Werner;R. Zawadzki
Single spectrometer-based complex conjugate artifact removal methods are evaluated for in vivo imaging with complementary metal-oxide semiconductor line scan camera based high-speed Fourier-domain optical coherence tomography (FD-OCT) at 100,000 axial scans per second. Performance of three different phase-shifting methods with the same OCT engine is evaluated using modified data acquisition schemes, depending on the requirements of each phase-shifting technique. The suppression ratio of complex conjugate artifact images using a paperboard is assessed for all tested methods. Several other characteristics, including a list of additional hardware requirements (beyond standard FD-OCT components) and data acquisition schemes for each of the methods is presented. In vivo full-range images of human fingerpad and nail are shown and compared with standard FD-OCT images. Additionally, a complex-conjugate-free human retinal volume acquired at the speed of 100,000 A-scans/s is presented.