Eluting trace components before major constituents. I: Sensitivity enhancement in analytical determinations of optical purity
Eluting trace components before major constituents. I: Sensitivity enhancement in analytical determinations of optical purity
复制标题
DOI:
10.1016/s0021-9673(01)94410-3
复制
发表时间:
1987
影响因子:
4.1
通讯作者:
J. A. Perry;J. D. Rateike;Ted Szczerba
中科院分区:
文献类型:
--
作者:
J. A. Perry;J. D. Rateike;Ted Szczerba
High optical purity can be precisely determined by high-performance liquid chromatography if several conditions can be met. Derivatization, if required, must be non-chiral; the elution order of enantiomers must be reversible, to allow elution of the trace enantiomer before the major enantiomer; and the column must under large overload retain efficiency for the trace enantiomer. Also for enhancing sensitivity to traces, we have found that a given quantity of sample is better injected in larger-than-conventional volume and at less-than-conventional concentration. Combining these approaches allowed not only the determination of 99.9967% optical purity in one demonstrative example but also an apparent extensibility to considerably higher purities. The limits to that extensibility, not obvious, were not established.