Quantification and minimization of magnetic susceptibility artifacts on GRE images
Quantification and minimization of magnetic susceptibility artifacts on GRE images
复制标题
DOI:
10.1097/00004728-200011000-00024
复制
发表时间:
2000-11-01
影响因子:
1.3
通讯作者:
Pomper, MG
中科院分区:
文献类型:
--
作者:
Port, JD;Pomper, MG
Purpose: The purpose of this work was to determine the optimal imaging parameters far minimization of metallic susceptibility artifacts during gradient echo (GRE) imaging.Method: We performed GRE imaging of titanium screws in a nickel-doped agarose gel phantom, systematically varying several parameters to characterize and quantify susceptibility artifacts.Results: The greatest reduction in artifact size came from using a short TE; increasing the frequency matrix and decreasing the slice thickness also contributed substantially to reducing the artifact size. Whenever possible, implanted prostheses should be aligned with the main magnetic field to minimize artifact size. Parameters with negligible effect on artifact size included bandwidth, phase encode matrix, and field of view.Conclusion: Radiologists can easily adjust the above parameters in their imaging protocols to improve GRE image quality in patients with implanted metallic devices.