Short-term reproducibility of apparent diffusion coefficient estimated from diffusion-weighted MRI of the prostate

Short-term reproducibility of apparent diffusion coefficient estimated from diffusion-weighted MRI of the prostate
复制标题

DOI:
10.1007/s00261-015-0396-x
复制
发表时间:
2015-10-01
期刊:
影响因子:
--
通讯作者:
Oto, Aytekin
Oto, Aytekin
中科院分区:
医学3区
文献类型:
--
作者:
Sadinski, Meredith;Medved, Milica;Oto, Aytekin

文献摘要

被引文献

相似文献

The purpose of the study is to determine short-term reproducibility of apparent diffusion coefficient (ADC) estimated from diffusion-weighted magnetic resonance (DW-MR) imaging of the prostate.Fourteen patients with biopsy-proven prostate cancer were studied under an Institutional Review Board-approved protocol. Each patient underwent two, consecutive and identical DW-MR scans on a 3T system. ADC values were calculated from each scan and a deformable registration was performed to align corresponding images. The prostate and cancerous regions of interest (ROIs) were independently analyzed by two radiologists. The prostate volume was analyzed by sextant. Per-voxel absolute and relative percentage variations in ADC were compared between sextants. Per-voxel and per-ROI variations in ADC were calculated for cancerous ROIs.Per-voxel absolute difference in ADC in the prostate ranged from 0 to 1.60 x 10(-3) mm(2)/s (per-voxel relative difference 0% to 200%, mean 10.5%). Variation in ADC was largest in the posterior apex (0% to 200%, mean 11.6%). Difference in ADC variation between sextants was not statistically significant. Cancer ROIs' per-voxel variation in ADC ranged from 0.001 x 10(-3) to 0.841 x 10(-3) mm(2)/s (0% to 67.4%, mean 11.2%) and per-ROI variation ranged from 0 to 0.463 x 10(-3) mm(2)/s (mean 0.122 x 10(-3) mm(2)/s).Variation in ADC within the human prostate is reasonably small, and is on the order of 10%.