Ion Peak Narrowing by Applying Additional AC Voltage (Ripple Voltage) to FAIMS Extractor Electrode
Ion Peak Narrowing by Applying Additional AC Voltage (Ripple Voltage) to FAIMS Extractor Electrode
复制标题
DOI:
10.1016/j.jasms.2009.09.001
复制
发表时间:
2010-01-01
影响因子:
3.2
通讯作者:
Sheven, Dmitriy G.
中科院分区:
文献类型:
--
作者:
Pervukhin, Viktor V.;Sheven, Dmitriy G.
The use of a non-uniform electric field in a high-field asymmetric waveform ion mobility spectrometry (FAIMS) analyzer increases sensitivity but decreases resolution. The application of an additional AC voltage to the extractor electrode ("ripple" voltage, U(ripple)) can overcome this effect, which decreases the FAIMS peak width. In this approach, the diffusion ion loss remains minimal in the non-uniform electric field in the cylindrical part of the device, and all ion losses under Uripple occur in a short portion of their path. Application of the ripple voltage to the extractor electrode is twice as efficient as the applying of U(ripple) along the total length of the device. (J Am Soc Mass Spectrom 2010, 21, 47-52) (C) 2010 American Society for Mass Spectrometry