Measurement of BOD: Correlation Between 5-Day BOD and Commercial BOD Biosensor Values
Measurement of BOD: Correlation Between 5-Day BOD and Commercial BOD Biosensor Values
复制标题
BOD 测量:5 天 BOD 与商业 BOD 生物传感器值之间的相关性
DOI:
10.1080/09593331808616544
复制
发表时间:
1997
期刊:
影响因子:
--
通讯作者:
Y. Asano
中科院分区:
文献类型:
--
作者:
J. Marty;D. Olive;Y. Asano
A flow injection system incorporating a commercial BOD sensor, using entrapped Trichosporum cutaneum, for the determination of BOD (biochemical oxygen demand) is presented. The linear range of BOD measurement varied from 10 to 70 mg l−1. The correlation between the BOD sensor and the official method (AFNOR T90 103 - France) is good for waste water inlet and outlet effluents and domestic landfill leachates up to 500 mg l−1. The lifetime of the membrane sensor is greater than 40 days and 200 measurements. Analysis time is 30 minutes including rinsing.