T1 AND T2 MEASUREMENTS ON A 1.5-T COMMERCIAL MR IMAGER

T1 AND T2 MEASUREMENTS ON A 1.5-T COMMERCIAL MR IMAGER
复制标题

DOI:
10.1148/radiology.171.1.2928538
复制
发表时间:
1989-04-01
期刊:
影响因子:
19.7
通讯作者:
HAUGHTON, VM
HAUGHTON, VM
中科院分区:
医学1区
文献类型:
--
作者:
BREGER, RK;RIMM, AA;HAUGHTON, VM

文献摘要

被引文献

相似文献

为了将弛豫时间用于临床诊断,必须确定测量的精确度。作者在体模和人类志愿者中测量了T1、T2和质子密度,以确定该方法的重复性。在两次软件升级的15个月期间,模体中T1测量的变异系数为5%。任何给定软件的T2测量的方差都在4%或更低,在15个月的总体期间,如果有两个软件更改,T2的重复性在6%到9%之间。重复性足够高,临床上精确测量T1、T2和质子密度是可行的。
In order for relaxation times to be used in clinical diagnosis, the precision of the measurement must be determined. The authors measured T1, T2, and proton density in a phantom and in human volunteers to determine the reproducibility of the method. The coefficient of variance of T1 measurements in the phantom during a 15-month period with two software upgrades was 5%. Variance of T2 measurements with any given software was 4% or less, and overall in the 15-month period, with two software changes, the T2 reproducibility was between 6% and 9%. The reproducibility is sufficiently high that precise clinical measurements of T1, T2, and proton density are feasible.