Digital image capture for high-resolution medical x-ray diagnostics
Digital image capture for high-resolution medical x-ray diagnostics
复制标题
用于高分辨率医学 X 射线诊断的数字图像捕获
DOI:
10.1117/12.2321166
复制
发表时间:
2018
期刊:
影响因子:
--
通讯作者:
Akif, Adil
中科院分区:
文献类型:
--
作者:
Mostovych, Nicholas A.;Morse, Theodore F.;Gupta, Rajiv;Murphy, Timothy P.;Weber, Peter M.;Cherepy, Nerine J.;Adams, Bernhard W.;Bifano, Thomas G.;Stankus, Brian;Akif, Adil
We have demonstrated a high resolution (10 micron) X-ray scintillator plate as part of an indirect X-ray detection system. Scintillator plates are typically integrated with a 2-dimensional array of photodiodes based upon amorphous Si. This paper describes an alternative digital capture system that leverages low cost CCD/CMOS cameras. Our detector has a broad set of potential applications, however the initial target application is mammography. Full-field mammography mandates an imaging area of 180mm x 240mm or larger. Very large CCD/CMOS sensors have recently been developed for high resolution cameras, such as the 250-pixel Canon camera which has sensor dimensions of 202mm x 205mm, and could conceivably be matched to our high-resolution scintillator plate without any intervening optics for magnification. However, such large format CCD/CMOS sensors have limited availability because of low production yields and high cost considerations. On the other hand, small form (36mm x 24mm) and medium format (44mm x 33mm) CCD/CMOS-based photodiodes have become widely available at low cost due to their applications in the large markets of mobile devices and consumer cameras. We have therefore developed a simple optical scheme for utilizing four small or medium format CCD/CMOS cameras to capture a larger, high-resolution image. Current systems employed in screening mammography resolve tissue features of 75-100 microns. Suspicious features found during preliminary mammographic screenings are further investigated during diagnostic mammographic tests which use a high-resolution detector that is focused over the suspicious lesion. Typically, an area less than 100mm x 80mm, the current maximum size of our high-resolution scintillation plate, is interrogated. We show that diagnostic mammography, over an area of 100mm x 80mm, could be performed using our system with a feature resolution down to 7 microns.
影响因子:
35
作者:
Yakunin S;Sytnyk M;Kriegner D;Shrestha S;Richter M;Matt GJ;Azimi H;Brabec CJ;Stangl J;Kovalenko MV;Heiss W
通讯作者:
Heiss W
DOI:
--
发表时间:
2012
期刊:
Nuclear Science Symposium and Medical Imaging Conference
影响因子:
--
作者:
N. Cherepy;R. Sanner;T. Tillotson;S. Payne;P. Beck;S. Hunter;L. Ahle;P. Thelin
通讯作者:
P. Thelin
DOI:
--
发表时间:
2012
期刊:
影响因子:
--
作者:
B. L. Rupert;N. Cherepy;B. Sturm;R. Sanner;S. Payne
通讯作者:
S. Payne