Investigation on the use of UTEVA as a stationary phase for chromatographic separation of actinides on-line to inductively coupled plasma mass spectrometry
Investigation on the use of UTEVA as a stationary phase for chromatographic separation of actinides on-line to inductively coupled plasma mass spectrometry
复制标题
使用 UTEVA 作为电感耦合等离子体质谱在线色谱分离锕系元素固定相的研究
DOI:
10.1039/b007123o
复制
发表时间:
2001
影响因子:
3.4
通讯作者:
R. Fuoco
中科院分区:
文献类型:
--
作者:
L. Perna;M. Betti;J. M. B. Moreno;R. Fuoco
A method based on the use of the UTEVA-Spec. Eichrom resin as the stationary chromatographic phase is described. The column was coupled on-line to an ICP-MS detector for the determination of actinides. The method is compared with one based on the use of commercially available cation-exchange stationary phases. The analytical procedure was validated by the use of certified reference materials as well as by other independent analytical techniques. It was demonstrated that the UTEVA-Spec. column can be used at the same time as a pre-concentration column, as well as a chromatographic column, lowering the detection limit of analysis down to a few pg g−1.