Stability control method of double crystal monochromator based on hybrid control algorithm
Stability control method of double crystal monochromator based on hybrid control algorithm
复制标题
DOI:
10.1016/j.nima.2022.167394
复制
发表时间:
2022-08
期刊:
影响因子:
--
通讯作者:
Yang Bai;Xuepeng Gong;Qipeng Lu;Yuan Song;Wanqian Zhu;S. Xue;Dazhuang Wang;Zhong-qi Peng;Zhen Zhang
中科院分区:
文献类型:
--
作者:
Yang Bai;Xuepeng Gong;Qipeng Lu;Yuan Song;Wanqian Zhu;S. Xue;Dazhuang Wang;Zhong-qi Peng;Zhen Zhang
The Bragg angle stability of the DCM (double crystal monochromator) in synchrotron radiation facilities is critical for beam position, flux and resolution of hard X-ray beam lines. Generally, the vibration caused by LN2cooling system is the most critical factor affecting the stability of DCM. In this paper, aiming at the stability of DCM, we propose the stability control method of double crystal monochromator base on hybrid control algorithm. The numerical results show that the stability of the DCM in the pitch direction can be greatly improved by this active vibration control method under LN2mode. The angular displacement RMS value in the pitch direction decreased from about 358.52 nrad to about 0.071nrad, which decreased by 99.98%. Proposed method is of great significance to the research of ultra-high stability DCM in synchrotron radiation facilities all over the world.