[Nitrous oxide induced intraluminal tube obstruction during endotracheal intubation with armored tubes (author's transl)].

[Nitrous oxide induced intraluminal tube obstruction during endotracheal intubation with armored tubes (author's transl)].
复制标题

[用铠装管进行气管插管期间一氧化二氮引起管腔内阻塞(作者翻译)]。

DOI:
--
复制
发表时间:
1981
期刊:
Der Anaesthesist
影响因子:
--
通讯作者:
H. Wehmer
H. Wehmer
中科院分区:
--
文献类型:
--
作者:
K. Kopp;H. Wehmer

文献摘要

被引文献

相似文献

正如病例报告所证明的那样,在麻醉期间使用笑气时,可能会发生铠装气管插管的危险阻塞。这是基于 N2O 扩散到管壁内部预先存在的气泡中,从而增加气泡体积。管壁中的气泡在生产和再灭菌过程中产生。避免对气管内铠装管进行重新消毒和仔细检查应可预防上述并发症。
As demonstrated by a case report, dangerous obstruction of armoured endotracheal tubes may occur when using nitrous oxide during anaesthesia. This is based on N2O diffusion into preexisting air bubbles of the inner part of the tube wall with consequent enhancement of the bubble volume. The bubbles in the tube wall arise during production as well as during resterilisation. Avoidance of resterilisation and meticulous examination of endotracheal armoured tubes should prevent the described complication.