ANALYSIS OF 6R-TETRAHYDROBIOPTERIN AND 6S-TETRAHYDROBIOPTERIN AND OTHER PTERINS BY REVERSED-PHASE ION-PAIR LIQUID-CHROMATOGRAPHY WITH FLUOROMETRIC DETECTION BY POSTCOLUMN SODIUM-NITRITE OXIDATION
ANALYSIS OF 6R-TETRAHYDROBIOPTERIN AND 6S-TETRAHYDROBIOPTERIN AND OTHER PTERINS BY REVERSED-PHASE ION-PAIR LIQUID-CHROMATOGRAPHY WITH FLUOROMETRIC DETECTION BY POSTCOLUMN SODIUM-NITRITE OXIDATION
复制标题
DOI:
10.1016/0378-4347(93)80495-p
复制
发表时间:
1993-08-11
期刊:
影响因子:
--
通讯作者:
OHNO, T
中科院分区:
文献类型:
--
作者:
TANI, Y;OHNO, T
A rapid and sensitive reversed-phase ion-pair liquid chromatographic system with fluorimetric detection by post-column sodium nitrite oxidation was established for measuring six pterin compounds (6S-5,6,7,8-tetrahydrobiopterin, 6S-5,6,7,8-tetrahydrobiopterin, 7,8-dihydrobiopterin, biopterin, pterin and D-neopterin). The coefficients of variation for these pterins were 0.705-3.714%, and the minimum detectable amount was ca. 10-20 pg at a signal-to-noise ratio of 3. A linear detector response was also verified. The concentrations of the pterin compounds in rat tissues were measured by the described method. Furthermore, by means of brain microdialysis, the output of pterin compounds from rat striatum was detected. Therefore, these results demonstrate that this system can be applied to analyses not only of various rat tissues but also of dialysates collected in vivo.