Differential mobility spectrometry of chlorocarbons with a micro-fabricated drift tube
Differential mobility spectrometry of chlorocarbons with a micro-fabricated drift tube
复制标题
DOI:
10.1039/b316326a
复制
发表时间:
2004-01-01
期刊:
影响因子:
4.2
通讯作者:
Miller, RA
中科院分区:
文献类型:
--
作者:
Eiceman, GA;Krylov, EV;Miller, RA
Chlorocarbons were ionized through gas phase chemistry at ambient pressure in air and resultant ions were characterized using a micro-fabricated drift tube with differential mobility spectrometry (DMS). Positive and negative product ions were characterized simultaneously in a single drift tube equipped with a 3 mCi Ni-63 ion source at 50 degreesC and drift gas of air with 1 ppm moisture. Scans of compensation voltage for most chlorocarbons produced differential mobility spectra with Cl- as the sole product ion and a few chlorocarbons produced adduct ions, M. Cl-. Detection limits were similar to 20-80 pg for gas chromatography-DMS measurements. Chlorocarbons also yielded positive ions through chemical ionization in air and differential mobility spectra showed peaks with characteristic compensation voltages for each substance. Field dependence of mobility was determined for positive and negative ions of each substance and confirmed characteristic behavior for each ion. A DMS analyzer with a membrane inlet was used to continuously monitor effluent from columns of bentonite or synthetic silica beads to determine breakthrough volumes of individual chlorocarbons. These findings suggest a potential of DMS for monitoring subsurface environments either on site or perhaps in situ.