ANTERIOR CRUCIATE LIGAMENT TEAR - PROSPECTIVE EVALUATION OF DIAGNOSTIC-ACCURACY OF MIDDLE-FIELD-STRENGTH AND HIGH-FIELD-STRENGTH MR-IMAGING AT 1.5 AND 0.5 T
ANTERIOR CRUCIATE LIGAMENT TEAR - PROSPECTIVE EVALUATION OF DIAGNOSTIC-ACCURACY OF MIDDLE-FIELD-STRENGTH AND HIGH-FIELD-STRENGTH MR-IMAGING AT 1.5 AND 0.5 T
复制标题
DOI:
10.1148/radiology.197.3.7480763
复制
发表时间:
1995-12-01
期刊:
影响因子:
19.7
通讯作者:
AMENDOLA, A
中科院分区:
文献类型:
--
作者:
VELLET, AD;LEE, DH;AMENDOLA, A
PURPOSE: To compare the diagnostic efficacy of current-generation middle- and high-field-strength magnetic resonance (MR) imagers in the diagnosis of anterior cruciate ligament (ACL) tears.MATERIALS AND METHODS: In 114 of 230 patients referred for knee imaging, MR imaging at 0.5 and 1.5 T was performed with identical sequences but with a slightly longer total imaging time and bandwidth optimization at 0.5 T. Radiologists were blinded to diagnosis and field strength. Sensitivity, specificity, and accuracy were determined, and ACL tear was confirmed by means of arthroscopy and pathology.RESULTS: There was no difference between the field strengths in accuracy, sensitivity, or specificity for the diagnosis of ACL tears in 86 patients with disrupted ACLs and 28 patients with intact ACLs. Accuracy for all ACL tears was 90% at 0.5 T and 91% at 1.5 T. Similarly, there were no differences in diagnosis of meniscal tears (79 with, 149 without) or posterior cruciate ligament tears (seven with, 107 without).CONCLUSION: Higher field strength does not confer higher accuracy in the diagnosis of ACL tears at MR imaging.