Noninvasive Imaging and Measurement of Accommodation Using Dual-Channel SD-OCT
Noninvasive Imaging and Measurement of Accommodation Using Dual-Channel SD-OCT
复制标题
使用双通道 SD-OCT 进行无创成像和调节测量
DOI:
10.3109/02713683.2013.860991
复制
发表时间:
2014-06-01
影响因子:
2
通讯作者:
Zhou, Chuanqing
中科院分区:
文献类型:
--
作者:
Sun, Yong;Fan, Shanhui;Zhou, Chuanqing
Purpose: To investigate the feasibility and practicality of real-time noninvasive imaging and measurement of ocular accommodation (0-5D with one diopter step) using dual-channel, dual-focus spectral domain optical coherence tomography (SD-OCT).Materials and Methods: A custom-built, dual-channel, dual-focus SD-OCT was used to image the anterior segment, including the cornea, iris, and anterior and posterior lens surfaces. Three consecutive measurements were taken in each accommodative session for each subject. Changes in ocular dimensions during accommodation were calculated based on the acquired SD-OCT images. A Friedman test was used to test sensitivity of ocular dimension changes per diopter of accommodation.Results: With accommodation, anterior chamber depth (ACD), curvature radius of both anterior (RAL) and posterior (RPL) lens surfaces, and pupil diameter (PD) decreased significantly, whereas lens thickness (LT) increased significantly (p