Intravascular ultrasound can be used to evaluate pulsed laser ablation of arterial tissue.

Intravascular ultrasound can be used to evaluate pulsed laser ablation of arterial tissue.
复制标题

血管内超声可用于评估动脉组织的脉冲激光消融。

DOI:
10.1002/lsm.1900160205
复制
发表时间:
1995
影响因子:
2.4
通讯作者:
Abela,GS
Abela,GS
中科院分区:
医学3区
文献类型:
--
作者:
Friedl,SE;Mathews,ED;Miyamoto,A;Abela,GS

文献摘要

相似文献

Background and Objective:Intravascular ultrasound (IVUS) has been used successfully to detect intravascular lesions. This study evaluates the ability of IVUS to detect acoustic damage to the arterial wall following high power, pulsed laser ablation.Study Design/Materials and Methods: Arterial ablation and disruption were performed in necropsy bovine aorta with a Ho:YAG laser using energy ranging from 140–720 mJ/pulse at 5 Hz. Laser energy was delivered with 2 mm diameter, multifiber over‐the‐wire catheters. A 20‐MHz IVUS catheter was used to image the arterial damage prior to tissue fixation and morphometry.Results:IVUS images revealed ablation craters surrounded by high acoustically backscattering zones. By histology, the arteries revealed ablation craters lined with thermal coagulation surrounded by a region of dissection and vacuolization. The depth and width of the highly backscattering zones on IVUS images correlated strongly with the corresponding morphometric measurements of tissue dissection (r = 0.92,P= 0.0001 and r = 0.80,P= 0.0001, respectively). Morphometric measurements of the ablation crater depth correlated strongly with laser energy (r = 0.90,P= 0.0001), whereas crater width was not correlated with laser energy (r = 0.27,P= 0.09).Conclusion:This study demonstrates that IVUS can detect and measure the extent of arterial damage following pulsed laser ablation. This may provide a means of detecting the extent of tissue disruption and help develop approaches to reduce or prevent extensive tissue damage. © 1995 Wiley‐Liss, Inc.