A measurement of the energy and timing resolution of the GlueX Forward Calorimeter using an electron beam
A measurement of the energy and timing resolution of the GlueX Forward Calorimeter using an electron beam
复制标题
使用电子束测量 GlueX 正向热量计的能量和时间分辨率
DOI:
10.1016/j.nima.2013.05.109
复制
发表时间:
2013
影响因子:
1.4
通讯作者:
H. Egiyan
中科院分区:
文献类型:
--
作者:
K. Moriya;J. Leckey;M. Shepherd;K. Bauer;D. Bennett;J. Frye;J. González;S. Henderson;D. Lawrence;R. Mitchell;E. Smith;Paul Smith;A. Somov;H. Egiyan
The performance of the GlueX Forward Calorimeter was studied using a small version of the detector and a variable energy electron beam derived from the Hall B tagger at Jefferson Lab. For electron energies from 110 MeV to 260 MeV, which are near the lower-limits of the design sensitivity, the fractional energy resolution was measured to range from 20% to 14%, which meets the design goals. The use of custom 250 MHz flash ADCs for readout allowed precise measurements of signal arrival times. The detector achieved timing resolutions of 0.38 ns for a single 100 mV pulse, which will allow timing discrimination of photon beam bunches and out-of-time background during the operation of the GlueX detector.