Cosine error for a class of hyperspectral irradiance sensors

Cosine error for a class of hyperspectral irradiance sensors
复制标题

DOI:
10.1088/0026-1394/50/3/187
复制
发表时间:
2013-06-01
期刊:
影响因子:
2.4
通讯作者:
Zibordi, G.
Zibordi, G.
中科院分区:
工程技术3区
文献类型:
--
作者:
Mekaoui, S.;Zibordi, G.

文献摘要

被引文献

相似文献

The cosine error of a class of in situ hyperspectral irradiance sensors largely applied for ocean colour investigations has been characterized for both in-air and in-water measurements. Results for in-air measurements indicate a slight wavelength dependence of the cosine error with differences up to 2% in the 412 nm to 865 nm spectral interval at 65 degrees zenith angle (i.e. the angle of incident irradiance with respect to the normal axis of the irradiance collector). However, the dependence of the cosine error on the zenith angle is generally quite marked and significantly varies from radiometer to radiometer with values ranging from -5% up to +7% at 65 degrees. Additionally, apart from the expected increase in the cosine error with the angle of incidence, in-air measurements generally indicate an irregular deviation from the ideal cosine response near the normal angle of incidence. A more pronounced increase in the cosine error is generally observed when radiometers are operated in water with respect to in air.