Multichannel microchip electrospray mass spectrometry
Multichannel microchip electrospray mass spectrometry
复制标题
DOI:
10.1021/ac9607119
复制
发表时间:
1997-02-01
影响因子:
7.4
通讯作者:
Karger, BL
中科院分区:
文献类型:
--
作者:
Xue, QF;Foret, F;Karger, BL
Microfabricated multiple-channel glass chips were successfully interfaced to an electrospray ionization mass spectrometer (ESI-RIS), The microchip device was fabricated by standard photolithographic, wet chemical etching, and thermal bonding procedures. A high voltage was applied individually from each buffer reservoir for spraying sample sequentially from each channel, With the sampling orifice of the MS grounded, it was found that a liquid flow of 100-200 nL/min was necessary to maintain a stable electrospray., The detection limit of the microchip MS experiment for myoglobin was found to be lower than 6 x 10(-8) M. Samples in 75% methanol were successfully analyzed with good sensitivity, as were aqueous samples. The parallel multiple-channel microchip system allowed ESI-MS analysis of different samples of standard peptides and proteins in one chip.