On-line coupling of macroporous resin column chromatography with direct analysis in real time mass spectrometry utilizing a surface flowing mode sample holder.

On-line coupling of macroporous resin column chromatography with direct analysis in real time mass spectrometry utilizing a surface flowing mode sample holder.
复制标题

DOI:
10.1016/j.aca.2013.12.014
复制
发表时间:
2014-02
影响因子:
6.2
通讯作者:
Shanshan Zeng;Lu Wang;Teng Chen;Hai-bin Qu
Shanshan Zeng;Lu Wang;Teng Chen;Hai-bin Qu
中科院分区:
化学1区
文献类型:
--
作者:
Shanshan Zeng;Lu Wang;Teng Chen;Hai-bin Qu

文献摘要

被引文献

相似文献

A surface flowing mode sample holder was designed as an alternative sampling strategy for direct analysis in real time mass spectrometry (DART-MS). With the sample holder, the on-line coupling of macroporous resin column chromatography with DART-MS was explored and the new system was employed to monitor the column chromatography elution process ofPanax notoginseng. The effluent from macroporous resin column was first diluted and mixed with a derivatization reagent on-line, and the mixture was then directly transferred into the ionization region of DART-MS by the sample holder. Notoginsenosides were methylated and ionized in a metastable helium gas stream, and was introduced into MS for detection. The on-line system showed reasonable repeatability with a relative standard deviation of 12.3% for the peak area. Three notoginsenosides, i.e. notoginsenoside R1, ginsenoside Rb1and ginsenoside Rg1, were simultaneously determined during the eluting process. The alteration of the chemical composition in the effluent was accurately identified in 9 min, agreeing well with the off-line analysis. The presented technique is more convenient compared to the traditional UPLC method. These results suggest that the surface flowing mode DART-MS has a good potential for the on-line process monitoring in the pharmaceutical industry.