Cycle 27 COS NUV Spectroscopic Sensitivity Monitor
Cycle 27 COS NUV Spectroscopic Sensitivity Monitor
复制标题
Cycle 27 COS NUV 光谱灵敏度监测仪
DOI:
--
复制
发表时间:
2019
期刊:
影响因子:
--
通讯作者:
B. James
中科院分区:
文献类型:
--
作者:
W. Fischer;B. James
Observations of HST spectrophotometric standard stars show that the COS NUV detector has a time-dependent sensitivity (TDS) that must be monitored and accounted for in flux calibration. Regular observations monitor the changes in sensitivity for three NUV gratings: G230L, G185M, and G225M. Because the sensitivity of the fourth grating, G285M, is now very low, it was removed from the routine monitoring program, and its use is discouraged for General Observer (GO) programs. Results from the Cycle 26 NUV TDS program show that the G230L and G185M gratings, which are coated in MgF2, exhibit trends consistent with little or no change. On the other hand, the G225M grating, which is bare aluminum, shows a significant sensitivity decline of −2.87% ± 0.18% yr−1. It was discovered during Cycle 26 that the current NUV TDSTAB overestimates the rate of sensitivity loss for each grating. NUV fluxes have therefore been slightly overcorrected by an amount that grows with time and have, in recent years, become inconsistent with the 5% accuracy specification. Work is underway to correct this.