A brush stimulator for functional brain imaging
A brush stimulator for functional brain imaging
复制标题
DOI:
10.1016/j.clinph.2007.08.024
复制
发表时间:
2007-12-01
影响因子:
4.7
通讯作者:
Hari, R.
中科院分区:
文献类型:
--
作者:
Jousmaki, V.;Nishitani, N.;Hari, R.
Objective: To describe a novel non-magnetic hand-held device to stimulate various parts of the skin and to evaluate its performance in magnetoencephalographic (MEG) recordings.Methods: The hand-held part of the device consists of an optic fiber bundle that forms a small brush. Half of the fibers emit modulated red light and the other half detect the reflected light from the skin so that the brush-to-skin contact is detected by means of reflectance.Results: Light tapping of the back of the hand at the innervation area of the radial nerve elicited clear responses in all 10 subjects studied, with the main deflections peaking 40-70 ms after the stimulus. The earliest responses, obtained with a higher number of averaged trials, peaked 27-28 ms after the tap to the left hand dorsum. Source analysis of the MEG signals indicated neuronal sources at the primary somatosensory (SI) cortex, with a clear somatotopical order for face vs. hand.Conclusions: The device seems feasible for both MEG and functional magnetic resonance imaging experiments to address functional anatomy of the human somatosensory system with a real-life like stimulation.Significance: Non-magnetic and artefact-free tactile stimulator with a selective stimulus offers new possibilities for experimental designs to study the human mechanoreceptor system. (c) 2007 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.