Quantitative assessment of atrial regional function using motion estimation computed tomography.
Quantitative assessment of atrial regional function using motion estimation computed tomography.
复制标题
DOI:
10.1097/rct.0000000000000108
复制
发表时间:
2014-09
影响因子:
1.3
通讯作者:
Taguchi K
中科院分区:
文献类型:
--
作者:
Ashikaga H;Cammin J;Tang Q;Knudsen K;Inoue Y;Fishman EK;Taguchi K
We validated a novel image-based motion estimation CT technique (iME) to quantify atrial regional function in swine in vivo. Domestic swine (n=8) underwent CT scan with intravenous contrast before and after median sternotomy where 15–30 glass beads were sutured to the atria to calculate the motion estimation error. Four-dimensional motion vector field (MVF) was estimated using iME. Area change ratio (%AC) was calculated over the atrial endocardium to assess the surface deformation. The error between the measured and the calculated coordinates based on MVF was 0.76±0.43 mm. %AC was regionally heterogeneous. %AC time course was significantly different between the right and the left atrium (P<0.001), and between the right atrial appendage and the right atrial chamber (P=0.004). Quantitative assessment of atrial regional function using iME is highly accurate. iME can quantify subtle regional dysfunction that is not apparent in global functional indices such as ejection fraction.