Comparison of modeling accuracy of amplitude distribution models for ultrasonic tissue characterization of liver fibrosis
Comparison of modeling accuracy of amplitude distribution models for ultrasonic tissue characterization of liver fibrosis
复制标题
DOI:
10.1109/ultsym.2016.7728528
复制
发表时间:
2016-09
期刊:
影响因子:
--
通讯作者:
Shohei Mori;S. Hirata;H. Hachiya;Tadashi Yamaguchi
中科院分区:
文献类型:
--
作者:
Shohei Mori;S. Hirata;H. Hachiya;Tadashi Yamaguchi
A quantitative diagnostic method for liver fibrosis using ultrasound echo signals is highly required. A probability density function (PDF) of echo envelope from a normal liver can be approximated by a Rayleigh distribution; however, the PDF of echo envelope from liver fibrosis deviates from the Rayleigh distribution. To evaluate tissue characteristics in the ultrasound B-mode image, several amplitude distribution models have been proposed. We proposed a multi-Rayleigh distribution model and evaluation method of liver fibrosis using the multi-Rayleigh model. In this study, we evaluated the modeling accuracy of the multi-Rayleigh model and other amplitude distribution models using the KL divergence. From the evaluated results for the 120 clinical data, it was found that the multi-Rayleigh model with three components is more suitable model than other amplitude distribution models for approximating the PDF of echo envelope from the liver fibrosis.