Physiologic responses to an MR imaging procedure performed at a specific absorption rate of 6.0 W/kg.

Physiologic responses to an MR imaging procedure performed at a specific absorption rate of 6.0 W/kg.
复制标题

对 6.0 W/kg 特定吸收率下进行的 MR 成像程序的生理反应。

DOI:
--
复制
发表时间:
1994
期刊:
影响因子:
19.7
通讯作者:
Emanuel Kanal
Emanuel Kanal
中科院分区:
医学1区
文献类型:
--
作者:
F. Shellock;D. Schaefer;Emanuel Kanal

文献摘要

被引文献

相似文献

目的
PURPOSE To assess the physiologic responses to a magnetic resonance (MR) procedure performed at a whole-body-averaged specific absorption rate (SAR) of 6.0 W/kg. MATERIALS AND METHODS Tympanic membrane temperature, skin temperature at seven sites, heart rate, blood pressure, oxygen saturation, and cutaneous blood flow in six volunteers were assessed before, for 16 minutes during, and immediately after MR imaging performed at 1.5 T, 64 MHz. RESULTS Statistically significant (P < .05) increases in temperature of the tympanic membrane and the skin of the chest, abdomen, upper arm, hand, and thigh and in heart rate and cutaneous blood flow were associated with exposure to high SAR. None of these physiologic changes, however, were considered deleterious to the study subjects. CONCLUSION MR imaging procedures performed at the high SAR studied can be tolerated by persons with normal thermoregulatory function.